Astra300™ Spirometer
TECHNICAL SPECIFICATIONS
Forced Parameters Measured | |
Forced vital capacity: | FVC, FEV.5, FEV1, FEV3, FEV6, FEV1/FVC, FEV1/FEV6, |
FEV.5/FVC, FEV3/FVC, PEF, FEF25, FEF50, FEF75, FEF25-75, | |
FEF 75-85, FET, FEV1/FEV.5, FEV1/PEF, MIF50, FIVC, FIV1, | |
FIV1/FVC, FEV1/FIV1,PIF, MTT, PEV/PIF, Lung Age | |
MVV: | Br/min |
Relaxed Parameters Measured | |
VC: | VC, TV, ERV, IRV, IC, Ti, Te, Tt, Ti/Tt |
Maneuvers per patient: | 5 total; best test and last test highlighted |
Interpretation: | Selectable – Kory, Miller, ATS, NLHEP (Ferguson) |
Predicted sets: | Selectable – 14 Total |
Ethnic correction: | Selectable – % Actual, Caucasian, Hispanic, AfroAmerican |
Language: | Selectable – English or Spanish |
Measurement Scale (BTPS) | |
Flow range | 0–16 L/sec |
Volume | 0–10 L |
Measurement method: | Infrared Interruption via Digital Turbine |
Dynamic flow resistance: | <1.5 cm H20/L/sec @ 14 L/sec |
Precision of Measurements (BTPS) | |
At highest volume: | +/- 3% or 50 mL |
At highest flow: | +/- 5% or 150 mL |
Time related precision: | .5% |
Volume resolution: | <6 mL |
Sampling frequency: | 50 Hz |
Hardware Specifications | |
Storage capacity: | 1000+ studies |
Keyboard: | Alphanumeric LCD touchscreen |
Communication: | USB to printer and PC, RS232 |
Optional: Bluetooth to PC | |
Dimensions: | 5.75" x 3.25" x 4.0" |
Power supply: | 2 x 1.5 DC Alkaline or NiMH batteries |
Weight: | 9 oz (250 g) |
Optional | OSHA module |
Flex Pulse Oximetry module and program | |
NiMH battery/charger pack | |
Calibration syringe | |
Astra300™ is compatible with printers using USB connection and PCL6 drivers. | |
|