| Forced Parameters Measured | 
  | 
| Forced vital capacity: | 
FVC, FEV1, FEV1/FVC, PEF, FEF50%,FEV6, FEV1/FEC, PEF, FEF50%, | 
|   | 
FEF25-75%, FEV6,FEV1/ FEV0.5, PEFT, Vext, FIVC, FIF50%, FEF50/FIF50, | 
|   | 
QC Grades, FEV.5, FEV3, FEV.5/FVC,FEV3/ FVC,FEV1/VC, FEV1/FEV6, | 
|   | 
FEV1/PEF, FEV1/FIV1, PEF/PIF, FEF25%, FEF75%, FEF75-85%, FET25-75%, | 
|   | 
FET100, FIV1, FIV1/FIVC, PIF, MTT, MVVInd, Lung age | 
|   | 
  | 
| MVV: | 
MVV, Br./min | 
|   | 
  | 
| Bronchoconstriction: | 
FVC, FEV1, PEF, FEF25-75%, PDx | 
|   | 
  | 
| Relaxed Parameters Measured | 
  | 
| VC: | 
VC, TV, ERV, IRV, IC, Ti, Te, Tt, Ti/Tt | 
|   | 
  | 
| Maneuvers per patient: | 
8 total; best test and last test highlighted | 
| Interpretation: | 
Selectable – Kory, Miller, ATS, NLHEP (Ferguson) | 
| Predicted sets: | 
Selectable – 16 Total | 
| Ethnic correction: | 
Selectable – % Actual, Caucasian, Hispanic, AfroAmerican | 
| Language: | 
Selectable – English, Spanish or French | 
|   | 
  | 
|   | 
  | 
| Measurement Scale (BTPS) | 
  | 
| Flow range | 
0–16 L/sec | 
| Volume | 
0–10 L | 
| Measurement method: | 
Infrared Interruption via Digital Turbine | 
|  Dynamic flow resistance: | 
<1.5 cm H20/L/sec @ 14 L/sec | 
|   | 
  | 
| Measurement Precision (BTPS) | 
  | 
| At highest volume: | 
+/- 3% or 50 mL | 
| At highest flow: | 
+/- 5% or 150 mL | 
| Time related precision: | 
.5% | 
| Volume resolution: | 
<6 mL | 
| Sampling frequency: | 
50 Hz | 
|   | 
  | 
| Hardware Specifications  | 
  | 
| Storage capacity: | 
3000+ study database | 
| Display: | 
5.7" High resolution color VGA touchscreen | 
|  Communication: | 
USB to printer and PC, RS232, HL-7* | 
|   | 
Optional: Bluetooth to PC | 
| Dimensions: | 
7.7" x 10.6"x 3.9" | 
| Power supply: | 
2 x 1.5 DC Alkaline or NiMH batteries | 
| Weight: | 
3.7 lbs | 
| Storage temperature: | 
-4 to 158 F | 
| Printer: | 
110 mm thermal graphic printer | 
|   | 
  | 
| Optional | 
Ethernet I/O | 
|   | 
Bluetooth to PC | 
|   | 
Flex Pulse Oximetry module and program (ask for indices) | 
|   | 
NiMH battery/charger pack | 
|   | 
Calibration syringe | 
|   | 
MIP/MEP | 
|   | 
Weather Station (Barometric Pressure, Temperature, Humidity) | 
|   | 
Sniff -- Rhinomanometry  | 
|   | 
Carrying Case | 
|   | 
  | 
| 
   
 |