Astra300™ Spirometer
TECHNICAL SPECIFICATIONS
Forced Parameters Measured | |
Forced vital capacity: | FVC, FEV.5, FEV1, FEV3, FEV6, FEV1/FVC, FEV1/FEV6, |
FEV.5/FVC, FEV3/FVC, PEF, FEF25, FEF50, FEF75, FEF25-75, | |
FEF 75-85, FET, FEV1/FEV.5, FEV1/PEF, MIF50, FIVC, FIV1, | |
FIV1/FVC, FEV1/FIV1,PIF, MTT, PEV/PIF, Lung Age | |
MVV: | Br/min |
Relaxed Parameters Measured | |
VC: | VC, TV, ERV, IRV, IC, Ti, Te, Tt, Ti/Tt |
Maneuvers per patient: | 5 total; best test and last test highlighted |
Interpretation: | Selectable – Kory, Miller, ATS, NLHEP (Ferguson) |
Predicted sets: | Selectable – 14 Total |
Ethnic correction: | Selectable – % Actual, Caucasian, Hispanic, AfroAmerican |
Language: | Selectable – English or Spanish |
Measurement Scale (BTPS) | |
Flow range | 0–16 L/sec |
Volume | 0–10 L |
Measurement method: | Infrared Interruption via Digital Turbine |
Dynamic flow resistance: | <1.5 cm H20/L/sec @ 14 L/sec |
Precision of Measurements (BTPS) | |
At highest volume: | +/- 3% or 50 mL |
At highest flow: | +/- 5% or 150 mL |
Time related precision: | .5% |
Volume resolution: | <6 mL |
Sampling frequency: | 50 Hz |
Hardware Specifications | |
Storage capacity: | 1000+ studies |
Keyboard: | Alphanumeric LCD touchscreen |
Communication: | USB to printer and PC, RS232 |
Optional: Bluetooth to PC | |
Dimensions: | 5.75" x 3.25" x 4.0" |
Power supply: | 2 x 1.5 DC Alkaline or NiMH batteries |
Weight: | 9 oz (250 g) |
Optional | OSHA module |
Flex Pulse Oximetry module and program | |
NiMH battery/charger pack | |
Calibration syringe | |
Astra300™ is compatible with printers using USB connection and PCL6 drivers. | |
|
AstraTouch™ Spirometer
TECHNICAL SPECIFICATIONS
Forced Parameters Measured | |
Forced vital capacity: | FVC, FEV1, FEV1/FVC, PEF, FEF50%,FEV6, FEV1/FEC, PEF, FEF50%, |
FEF25-75%, FEV6,FEV1/ FEV0.5, PEFT, Vext, FIVC, FIF50%, FEF50/FIF50, | |
QC Grades, FEV.5, FEV3, FEV.5/FVC,FEV3/ FVC,FEV1/VC, FEV1/FEV6, | |
FEV1/PEF, FEV1/FIV1, PEF/PIF, FEF25%, FEF75%, FEF75-85%, FET25-75%, | |
FET100, FIV1, FIV1/FIVC, PIF, MTT, MVVInd, Lung age | |
MVV: | MVV, Br./min |
Bronchoconstriction: | FVC, FEV1, PEF, FEF25-75%, PDx |
Relaxed Parameters Measured | |
VC: | VC, TV, ERV, IRV, IC, Ti, Te, Tt, Ti/Tt |
Maneuvers per patient: | 8 total; best test and last test highlighted |
Interpretation: | Selectable – Kory, Miller, ATS, NLHEP (Ferguson) |
Predicted sets: | Selectable – 16 Total |
Ethnic correction: | Selectable – % Actual, Caucasian, Hispanic, AfroAmerican |
Language: | Selectable – English, Spanish or French |
Measurement Scale (BTPS) | |
Flow range | 0–16 L/sec |
Volume | 0–10 L |
Measurement method: | Infrared Interruption via Digital Turbine |
Dynamic flow resistance: | <1.5 cm H20/L/sec @ 14 L/sec |
Measurement Precision (BTPS) | |
At highest volume: | +/- 3% or 50 mL |
At highest flow: | +/- 5% or 150 mL |
Time related precision: | .5% |
Volume resolution: | <6 mL |
Sampling frequency: | 50 Hz |
Hardware Specifications | |
Storage capacity: | 3000+ study database |
Display: | 5.7" High resolution color VGA touchscreen |
Communication: | USB to printer and PC, RS232, HL-7* |
Optional: Bluetooth to PC | |
Dimensions: | 7.7" x 10.6"x 3.9" |
Power supply: | 2 x 1.5 DC Alkaline or NiMH batteries |
Weight: | 3.7 lbs |
Storage temperature: | -4 to 158 F |
Printer: | 110 mm thermal graphic printer |
Optional | Ethernet I/O |
Bluetooth to PC | |
Flex Pulse Oximetry module and program (ask for indices) | |
NiMH battery/charger pack | |
Calibration syringe | |
MIP/MEP | |
Weather Station (Barometric Pressure, Temperature, Humidity) | |
Sniff -- Rhinomanometry | |
Carrying Case | |
|
AstraPulse FT™ Oximeter
TECHNICAL SPECIFICATIONS
Patient Range | Adult, Pediatric |
SpO2 | |
Range: | 35-99% |
Resolution: | 1% |
Accuracy: | 70% - 100% ±2% |
0% - 90% - unspecified | |
Heart (Pulse) Rate | |
Range: | 30-240 bpm |
Resolution: | 1 bpm |
Accuracy: | ± 2 bpm or ±2% |
Display | |
Type: | B&W Organic LED |
Parameters Displayed: | Sp02, PR, Sp02 waveform (pleth or line), signal strength bargraph |
Information: | Low battery |
Brightness: | 10 Levels |
Display Modes: | Six xelectable |
Environmental | |
Operating Temperature: | 5°C - 40°C |
Storage Temperature: | -10°C - 40°C |
Operating Humidity: | 15% RH - 80% RH |
Storage Humidity: | 10% RH - 80% RH |
Mechanical | |
Size: | 50 x 32 x 34 mm (LxWxH) |
Weight: | 50g (including batteries) |
Power: | 2 AAA alkaline or rechargeable NiMH batteries |
Operating Time: | Typical 30 hours |
Alarm | |
Finger out, low power | |
Auto power off - 10 seconds | |
|
AstraPulse FTP™ Oximeter
TECHNICAL SPECIFICATIONS
SpO2 | |
Range: | 70-100% |
Resolution: | 1% |
Accuracy: | 80% - 99% ±2% |
70% - 80% ±3% | |
Heart (Pulse) Rate | |
Range: | 30-235 bpm |
Resolution: | 1 bpm |
Accuracy: | ± 2 bpm or ±2% |
Audible: | PR tone modulation |
Display | |
Type: | TruColor Organic LED |
Parameters Displayed: | Sp02, PR, Sp02 waveform (pleth or line), signal strength bargraph |
Information: | Low battery |
Brightness: | 10 Levels |
Display Modes: | Six xelectable |
Environmental | |
Operating Temperature: | 5°C - 40°C |
Storage Temperature: | -10°C - 40°C |
Operating Humidity: | 15% RH - 80% RH |
Storage Humidity: | 10% RH - 80% RH |
Mechanical | |
Size: | 50 x 28 x 28 mm (LxWxH) |
Weight: | 1 .76 oz. (including batteries) |
Power: | 2 AAA alkaline or rechargeable NiMH batteries |
Operating Time: | Typical 20 hours |
|
PulmoGuard™
TECHNICAL SPECIFICATIONS
Pulmoguard fits the following instruments without an adapter:
Adapters available for instruments not listed.
Note: Use Pulmoguard II for all SDI models.
Assess Peak Flow Meter
Burdick - Spirosense
Circadian - Models (if using black Tamarac pneumotach, need 7975)
Cybermedics -Spinnaker TL, SL, CM3, CM310 (if white or black cone use 7980)
Cybermedics - Moose
Godart - All Models
Gould - 5004
Healthscan - Assess Peak Flow Meter
Jaeger - All Models except Spiro Pro
Jones - All Models except Satellite (small end of filter)
Med Graphics - 1070, 1085 (w/Prevent, use PG II; if no Prevent, 7970-spirometry, 7961-vol/diff)
Med Science - 570, 6257
Med Science - 3000, 4000 (spirometry-no adapt./vol-7961/diff-7960B)
Med Systems - Spirolite 101 & 201 (small end of filter), 343
New Med - 7000 I (Spirobyte)
PK Morgan - Benchmark (spirometry-7978, vol&diff-7960B), Flexi Flow (small end of filter), Trans Flow
Pneumedics - All Models (small end of filter)
Riko - AS300
S&M - CF-150B, GS520, Keystone III, Portascreen, Pulmoscreen
ViaSys/Sensormedics - Horizon, Ohio, 922, 2130, 2200, 2400, 2450, 2800, 6200, 2100 (spirometry-7961, vol-7960B, diff-7960B), Vmax (can use PGII w/7976 or 7977)
Spirolink - All Models
Spirometrics - LTE, 2500 LTE
Spiroscreen - All Models (small end of filter)
Spirotech - All Models
Tamarac - All Models except Presto Flash